REG-ISO14971

Integrate ISO 14971:2019 device risk management into software development: hazard identification, risk evaluation, control implementation, verification, and residual risk documentation. For IEC 62304 Clause 7 software-specific risk activities, also load `REG-IEC62304-SWRM`.

v1.1.0New

Unsigned, install at your own risk

Unverified

This skill has no cryptographic signature attached. We can't verify the contents match what the publisher intended.

Install this skill

No account required. Pick your tool, or use the command that detects it for you.

Run in your terminal. It detects your AI tool.

npx @skills-hub-ai/cli install meddev-skills-reg-iso14971
Or get the file:
Browse all CLI commands →
Manual setup by platform

Claude Code

~/.claude/skills/meddev-skills-reg-iso14971/SKILL.md

Setup guide →
0Save

Instructions

ISO 14971 Risk Management for Software

Purpose

Integrate ISO 14971:2019 device risk management into software development: hazard identification, risk evaluation, control implementation, verification, and residual risk documentation. For IEC 62304 Clause 7 software-specific risk activities, also load REG-IEC62304-SWRM.

When to Apply

  • Any change affecting hazards, controls, or mitigations.
  • New features, interfaces, or protocols.
  • Safety classification updates or architecture changes.
  • Cybersecurity-relevant changes (maps into threat modeling skill if used).

Requirements (testable)

  1. Hazard Identification: Maintain a software hazard list with causes, sequences, and harms per ISO 14971:2019 5–6. Rationale: foundation for controls.
  2. Risk Evaluation: Use defined criteria for severity/probability; classify risks and document rationale. Rationale: consistent decisions.
  3. Risk Control Implementation: For each unacceptable risk, implement controls in design/code with explicit link to RISK-CTRL-### per 7. Rationale: traceable controls.
  4. Verification of Controls: Each control has verification evidence (test, analysis, inspection) showing risk reduction per 8. Rationale: effectiveness.
  5. Residual Risk: Record residual risk acceptability and justification; escalate if aggregate risk changes per 8. Rationale: decision visibility.
  6. Safe State Definition: Define and implement safe states for hazardous conditions; test transitions. Rationale: bounded failure behavior.
  7. Fault Handling: Link fault handlers to hazards/controls and verify logging/annunciation where applicable. Rationale: detect/respond to faults.
  8. Documentation: Maintain risk management file artifacts updated with changes (hazards, controls, verification, residuals). Rationale: auditability.
  9. Commit Traceability: Commits modifying controls or hazards must include IDs (RISK-###, RISK-CTRL-###, REQ-###). Rationale: change traceability.

Recommended Practices

  • Keep a single source of truth: risk-register.yaml with hazards, controls, verification.
  • Use templated commit messages: RISK: RISK-12 mitigated via RISK-CTRL-7; TEST-305 added.
  • Align safe-state logic with architecture segregation (see safety-classification skill).
  • Pair risk controls with monitoring/diagnostics to detect degradation.

Patterns

Risk control with traceability and safe state:

// HZ-07: Over-infusion; RISK-CTRL-19: clamp and alarm within 100 ms
// REQ-221, TEST-410 covers timing
bool enforce_over_infusion_limit(float rate_ml_per_hr) {
    if (rate_ml_per_hr > MAX_RATE_ML_PER_HR) {
        clamp_line();            // physical mitigation
        alarm_over_rate();       // user mitigation
        enter_safe_state();      // safe state: pump stopped, alarm latched
        return false;
    }
    return true;
}

Residual risk documentation (YAML):

hazard: HZ-07
control: RISK-CTRL-19
verification:
  - TEST-410: timing <100 ms pass
  - TEST-411: alarm annunciation pass
residual_risk: acceptable_per_policy_RP-02
justification: Meets severity/probability target after control per table RM-1

Commit message format:

feat: enforce over-infusion limit

RISK: HZ-07 mitigated by RISK-CTRL-19
REQ: REQ-221 updated
TEST: TEST-410 added, TEST-411 updated

Anti-Patterns (risks)

  • Generic “improved safety” commits without IDs -> risk: lost traceability.
  • Controls implemented without verification -> risk: unproven effectiveness.
  • Safe state undefined or unreachable -> risk: uncontrolled hazardous behavior.
  • Fault handlers that log only -> risk: no actual mitigation.
  • Residual risk not reassessed after change -> risk: aggregate risk drift.

Verification Checklist

  • Hazards updated for new/changed functionality; causes and harms documented.
  • Risk evaluation uses defined criteria; severity/probability recorded.
  • Controls mapped to hazards with IDs (RISK-CTRL-###) and implemented.
  • Verification evidence exists and passes for each control.
  • Safe states defined, reachable, and tested for hazardous conditions.
  • Fault handling paths tied to hazards; logging/annunciation verified.
  • Residual risk acceptability documented; escalation performed if needed.
  • Commits touching hazards/controls include IDs in message.
  • Risk management file artifacts regenerated/baselined after change.

Traceability

  • Use stable IDs: hazards (HZ-###), controls (RISK-CTRL-###), requirements (REQ-###), tests (TEST-###).
  • Embed IDs in code comments, test names, and commit messages.
  • Maintain risk-register.yaml or equivalent; generate hazard->control->verification views.

References

  • ISO 14971:2019, clauses 4–10 (risk management process).
  • REG-IEC62304-SWRM (IEC 62304:2006+A1:2015, Clause 7 — software risk process).
  • IEC TR 80002-1:2019 (guidance on applying ISO 14971 to software; second edition).
  • FDA Guidance: "Content of Premarket Submissions for Device Software Functions" (2023) for submission expectations.
  • FDA Guidance: "Cybersecurity in Medical Devices" (2023) for security risk management alignment.
  • MDCG 2019-16 (guidance on cybersecurity for medical devices; EU MDR context).

Changelog

  • 1.1.0 (2026-05-21): Clarified scope vs REG-IEC62304-SWRM (Clause 7).
  • 1.0.1 (2026-01-04): Audit corrections - updated IEC TR 80002-1 to 2019 edition, clarified FDA guidance references.
  • 1.0.0 (2026-01-04): Initial skill with hazard/control patterns, safe state guidance, commit traceability, and verification checklist.

Audit History

  • 2026-01-04: Regulatory audit performed. Corrections applied:
    • IEC TR 80002-1:2009 updated to 2019 edition (second edition)
    • Generic "FDA premarket software guidance" replaced with specific 2023 guidance titles
This skill doesn’t include stateful context yet, instructions only. Learn about stateful skills.

Security

Loading security scan...

Reviews (0)

Browse all →
meddev-authenticationDefine secure user/device authentication, session management, and RBAC suitable for medical device risk profiles. Use when working on authentication in medical device software.0 installsmeddev-state-machinesDefine explicit, testable state machines with safe states, controlled transitions, and error handling suited to medical device safety classes. Use when working on state machines in medical device software.0 installsmeddev-change-controlDefine controlled change-request workflow for software: impact analysis, approvals, verification scope, and regulatory submission flags. Configuration identification and baselines are in DOC-SCM (IEC 62304 Clause 8.1, 8.3); approved implementation and re-verification align with 8.2. Use when working on change control in medical device software.0 installsCONN-USBImplement safe USB device connectivity: class selection, enumeration handling, data validation, electrical and safety considerations.0 installsmeddev-ai-ml-medical-device-softwareAddress machine learning inside the regulated product (not LLM coding assistants): Good Machine Learning Practice (GMLP), training/validation data integrity, Predetermined Change Control Plans (PCCP) where applicable, drift monitoring, and linking models to ISO 14971 and software lifecycle controls. Use when working on ai ml medical device software in medical device software.0 installsTEST-HILValidate software with real or representative hardware, sensors, actuators, and timing, to uncover integration and safety issues not visible in simulation.0 installs

More from Medical Device Skills

View source →

More Build skills

Browse category →
senior-frontendBuild and review React, Next.js, TypeScript or Tailwind frontends: components, scaffolding, performance and bundle size. Use whenever the user wants a new component or app, a page feels slow, the bundle is too big, or asks for a frontend code review.643 installsui-design-systemUse when building a design system: tokens, component docs, responsive scales or developer handoff. Generates tokens with design_token_generator.py.546 installsfrontend-designGuidance for distinctive, intentional visual design when building new UI or reshaping an existing one. Helps with aesthetic direction, typography, and making choices that don't read as templated defaults.354 installssenior-backendBuild and fix backend services: REST or GraphQL APIs, auth, microservices, PostgreSQL queries, migrations and load tests. Use whenever the user's API is slow or broken, needs login or auth added, a schema change, or a new endpoint or service scaffolded.253 installsself-improving-agentCurate what the agent has learned into durable project knowledge. Reviews Claude Code auto-memory (MEMORY.md and its topic files) for learnings worth keeping, promotes proven ones into CLAUDE.md or .claude/rules/, records a correction as a lesson that persists across sessions, prunes stale or duplicate entries, and turns a recurring fix into a reusable skill. Use whenever the user asks what you have learned or saved about the project, which memories should become permanent rules, to remember a lesson or stop repeating a mistake, or to clean up, audit or check memory health.180 installsusing-superpowersUse when starting any conversation - establishes how to find and use skills, requiring skill invocation before ANY response including clarifying questions146 installs

Frequently asked questions about REG-ISO14971

What does the REG-ISO14971 skill do?

Integrate ISO 14971:2019 device risk management into software development: hazard identification, risk evaluation, control implementation, verification, and residual risk documentation. For IEC 62304 Clause 7 software-specific risk activities, also load `REG-IEC62304-SWRM`. It's a reusable SKILL.md instruction set that loads into your AI coding assistant on demand, no prompt engineering, no copy-pasting every session.

How do I install the REG-ISO14971 skill?

Run `npx @skills-hub-ai/cli install meddev-skills-reg-iso14971` from your terminal. The CLI writes the SKILL.md to the correct location for your AI tool (e.g. ~/.claude/skills/meddev-skills-reg-iso14971/ for Claude Code or ~/.cursor/skills/ for Cursor with --target cursor) and adds it to your project's .skills.json lockfile.

Which AI tools does REG-ISO14971 work with?

REG-ISO14971 runs in Claude Code. It follows the open Agent Skills standard (SKILL.md), so the same skill works in every supported tool without modification.

Is the REG-ISO14971 skill free?

Yes. Every skill on skills-hub.ai is free and open-source. There are no premium tiers, paywalls, or usage limits. You only pay for whatever AI assistant you're already using.

How do I use REG-ISO14971 after installing it?

In Claude Code, type `/meddev-skills-reg-iso14971` (or whatever slash command the skill registers) and the AI follows the skill's instructions immediately. You can also reference it by name in natural language, your AI loads the skill into context when relevant.

Can I share the REG-ISO14971 skill with my team?

Yes. Commit your project's .skills.json lockfile and teammates run `npx @skills-hub-ai/cli install` (no args) to install every skill at the exact version you pinned. Organization-scoped installs work via skills-hub.ai organizations.