Instructions
Unit Testing for Medical Device Software
Purpose
Establish patterns for unit tests that verify software units against requirements with safety-class-appropriate rigor.
When to Apply
- New or changed functions/modules, especially safety-related.
- Bug fixes to prevent regression.
Requirements (testable)
- Implement Units [Class A, B, C]: Implement each software unit per design. Rationale: 5.5.1.
- Verification Process [Class B, C]: Define strategies, methods, and procedures for unit verification; evaluate test procedure adequacy. Rationale: 5.5.2.
- Acceptance Criteria [Class B, C]: Define acceptance criteria before integration; confirm units meet criteria (requirements, interface consistency, coding standards). Rationale: 5.5.3.
- Class C Criteria [Class C]: Where design requires, add criteria for event sequencing, data/control flow, resource use, fault handling, initialization, diagnostics, memory behaviour, and boundary conditions. Rationale: 5.5.4 (engineering checklist).
- Execute and Record [Class B, C]: Perform unit verification and document results. Rationale: 5.5.5.
- Structure: Use Arrange-Act-Assert; deterministic tests with clear preconditions. Rationale: reliable evidence.
- Isolation: Mock hardware/HAL; no real hardware in unit tests. Rationale: determinism.
- Traceability: Each test links to requirement and risk control IDs. Rationale: compliance.
- CI Integration: Run in CI; archive results. Rationale: continuous assurance.
Recommended Practices
- Keep tests fast (<1s each) to encourage frequent runs.
- Use table-driven tests for combinations; parametrize to reduce duplication.
- Enforce
-Wall -Wextra -Werrorin test builds. - Add sanitizers (ASan/UBSan) on host where possible for non-embedded logic.
Patterns
C unit test with Unity (example):
// TEST-UNIT-105 covers REQ-C-TYPE-01
TEST(SumSamples, AccumulatesAll) {
uint16_t s[] = {1, 2, 3};
CHECK_EQUAL(6, sum_samples(s, 3));
}
Boundary test:
// TEST-UNIT-110 covers REQ-DEF-INPUT-01
TEST(ValidateFlow, RejectsNegative) {
CHECK_FALSE(validate_flow(-1.0f));
}
Mock HAL:
// TEST-UNIT-120 covers REQ-HAL-IF-01
MOCK_FUNCTION_WITH_RETURN(hal_status_t, hal_uart_write, const uint8_t*, size_t, size_t*);
Anti-Patterns (risks)
- Tests that depend on real timing or hardware -> risk: flakiness.
- No negative/boundary tests -> risk: missed edge defects.
- Untagged tests with requirements -> risk: broken traceability.
- Ignoring test failures in CI -> risk: regressions ship.
Verification Checklist
- Tests follow AAA and are deterministic.
- Coverage meets class targets (branch/MC-DC for critical logic).
- Hardware/HAL calls mocked; no real hardware dependencies.
- Requirements/risk IDs referenced in tests.
- Negative/boundary cases included.
- Tests run in CI with warnings-as-errors; failures block merges.
Traceability
- Map
TEST-UNIT-###toREQ-###; store coverage reports per release.
References
- IEC 62304:2006+A1:2015, 5.5 (software unit implementation and verification).
TEST-COVERAGE(informative MC/DC depth by class).- ISO 14971:2019 for risk-based test depth.
Changelog
- 1.1.0 (2026-05-21): Aligned with 5.5.2–5.5.5 and Class C acceptance topics.
- 1.0.0 (2026-01-04): Initial unit testing skill with coverage targets and traceability.
Audit History
- 2026-01-04: Audit performed. Verified:
- IEC 62304 section 5.5 reference for unit implementation/verification is accurate
- MC/DC requirement for Class B/C control logic aligns with industry practice
- Unity test framework examples are syntactically correct
Security
Reviews (0)
Related skills
Browse all →More from Medical Device Skills
View source →More Test skills
Browse category →Frequently asked questions about TEST-UNIT
What does the TEST-UNIT skill do?
Establish patterns for unit tests that verify software units against requirements with safety-class-appropriate rigor. It's a reusable SKILL.md instruction set that loads into your AI coding assistant on demand, no prompt engineering, no copy-pasting every session.
How do I install the TEST-UNIT skill?
Run `npx @skills-hub-ai/cli install meddev-skills-test-unit` from your terminal. The CLI writes the SKILL.md to the correct location for your AI tool (e.g. ~/.claude/skills/meddev-skills-test-unit/ for Claude Code or ~/.cursor/skills/ for Cursor with --target cursor) and adds it to your project's .skills.json lockfile.
Which AI tools does TEST-UNIT work with?
TEST-UNIT runs in Claude Code. It follows the open Agent Skills standard (SKILL.md), so the same skill works in every supported tool without modification.
Is the TEST-UNIT skill free?
Yes. Every skill on skills-hub.ai is free and open-source. There are no premium tiers, paywalls, or usage limits. You only pay for whatever AI assistant you're already using.
How do I use TEST-UNIT after installing it?
In Claude Code, type `/meddev-skills-test-unit` (or whatever slash command the skill registers) and the AI follows the skill's instructions immediately. You can also reference it by name in natural language, your AI loads the skill into context when relevant.
Can I share the TEST-UNIT skill with my team?
Yes. Commit your project's .skills.json lockfile and teammates run `npx @skills-hub-ai/cli install` (no args) to install every skill at the exact version you pinned. Organization-scoped installs work via skills-hub.ai organizations.