REG-FDA820

Map **21 CFR Part 820 Subpart C — Design Controls** to software lifecycle work: design planning, inputs, outputs, review, verification, validation, transfer, and DHF/DMR links, including how **software validation** (intended use / risk controls) complements **verification** (requirements and specs). Includes a concise **ISO 13485:2016** slice where it overlaps QMS expectations for design and risk (interpretation; align with your notified-body MDSAP scope for EU).

v1.0.0New

Unsigned, install at your own risk

Unverified

This skill has no cryptographic signature attached. We can't verify the contents match what the publisher intended.

Install this skill

No account required. Pick your tool, or use the command that detects it for you.

Run in your terminal. It detects your AI tool.

npx @skills-hub-ai/cli install meddev-skills-reg-fda820
Or get the file:
Browse all CLI commands →
Manual setup by platform

Claude Code

~/.claude/skills/meddev-skills-reg-fda820/SKILL.md

Setup guide →
0Save

Instructions

FDA Design Controls and Software Validation (21 CFR 820.30)

Purpose

Map 21 CFR Part 820 Subpart C — Design Controls to software lifecycle work: design planning, inputs, outputs, review, verification, validation, transfer, and DHF/DMR links, including how software validation (intended use / risk controls) complements verification (requirements and specs). Includes a concise ISO 13485:2016 slice where it overlaps QMS expectations for design and risk (interpretation; align with your notified-body MDSAP scope for EU).

When to Apply

  • Defining or updating design procedures, DHF artifacts, or release records for US market devices.
  • Interpreting “validation” vs “verification” for software in FDA QMS language.
  • Connecting IEC 62304 deliverables to 820.30 design control records.
  • Preparing for ISO 13485 audits that reference design control (Clause 7.3).

Requirements (testable)

  1. Design Plan: Establish and maintain a design and development plan per 820.30(b); scope, responsibilities, interfaces, and review gates; update when software scope or class changes. Rationale: controlled design process.
  2. Design Inputs: Capture physical and performance requirements from intended use and safety/risk; trace to 820.30(c) records. Software inputs include functional, safety, security, usability, and regulatory constraints. Rationale: correct inputs drive correct outputs.
  3. Design Outputs: Document 820.30(d) outputs (specs, code baselines, build artifacts) that meet design inputs; include acceptance criteria for verification. Rationale: objective evidence of “what was built.”
  4. Design Review: Hold 820.30(e) reviews at planned stages; record participants, decisions, and action items for software architecture and major changes. Rationale: defect prevention.
  5. Design Verification: Demonstrate 820.30(f) that design outputs meet design inputs (tests, analysis, inspection). Map to unit/integration/system tests and static analysis. Rationale: “did we build it right?”
  6. Design Validation: Demonstrate 820.30(g) that approved device meets user needs and intended use under simulated or actual use as appropriate; include risk-related use scenarios. For software, often clinical/ summative usability and end-to-end clinical workflows where applicable. Rationale: “did we build the right thing?”
  7. Design Transfer: Ensure 820.30(h) transfer to manufacturing / deployment (including cloud release procedures) with verified procedures and training. Rationale: reproducible release.
  8. Design History File (DHF): Maintain 820.30(j) compilation of records for each design or type; link to SDD, tests, reviews, risk file, and change records. Rationale: audit trail.
  9. Device Master Record (DMR): Where applicable, ensure 820.181 DMR entries for released software (labels, IFU, build IDs) are consistent with validated design. Rationale: correct field configuration.
  10. ISO 13485 alignment (slice): Where QMS is ISO 13485-certified, map 7.3 Design and development and 4.1.6 (software used in QMS) to your procedures; document software validation for GMP/QMS tools per FDA Computer Software Assurance guidance when those tools affect production/quality records. Rationale: single coherent QMS story.

Recommended Practices

  • Maintain a one-page 820.30 ↔ IEC 62304 mapping table (SDP ↔ design plan, SOUP ↔ design input/output, etc.).
  • Separate verification matrix (REQ→test) from validation summary (intended use / risk / use scenarios).
  • Baseline software build IDs in DHF per release.

Patterns

DHF index entry (YAML):

dhf:
  device: "Pump-Controller-SW"
  design_plan_ref: "DOC-DSP-001 rev C"
  inputs_ref: ["REQ-SYS-001", "RMF-2026"]
  outputs_ref: ["SAD-001", "SDD-002", "tag: v2.3.1"]
  verification:
    summary: "reports/VV-REPORT-2.3.1.pdf"
    matrix: "trace/req-test-matrix.csv"
  validation:
    summary: "reports/VAL-SUMMARY-2.3.1.pdf"
    note: "Intended use + clinical workflow scenarios; links to IEC 62366 summative tests if UI"
  reviews:
    - id: "DR-2026-01"
      date: "2026-01-04"
      attendees: ["SE", "QA", "RA"]

Commit message tying to design controls:

feat: door interlock stop within 50 ms

820.30: design output + verification for REQ-62304-102
DHF: VV-REPORT updated; see TEST-201, TEST-202

Anti-Patterns (risks)

  • Calling integration tests “validation” only — risk: missing intended use evidence under 820.30(g).
  • DHF references only code without inputs/outputs linkage — risk: 483 observations on incomplete design controls.
  • Skipping design transfer for cloud/SaaS deployments — risk: production/tenant config drift.

Verification Checklist

  • Design plan current; reviews documented for major software milestones.
  • Design inputs traceable to risks and intended use; approved.
  • Design outputs (specs, builds) linked to inputs; acceptance criteria defined.
  • Verification evidence (tests, analysis) covers design outputs vs inputs.
  • Validation summary addresses user needs / intended use (not only unit tests).
  • Design transfer records exist for manufacturing/deployment.
  • DHF contains or references required records for the software release.
  • DMR/labeling consistent with released software version.

Traceability

  • Tag DHF artifacts with release vX.Y.Z; link REQ-* to TEST-* and validation VAL-* or SUM-UE-* (usability).
  • Map RISK-CTRL-* to both verification (technical) and validation (use-related) where applicable.

References

  • 21 CFR Part 820, Subpart C — Design controls (§820.30).
  • 21 CFR Part 820, Subpart M — Records (§820.181 DMR).
  • FDA Guidance: “General Principles of Software Validation” (2002) — validation philosophy (interpretation; not superseded for QMS concepts).
  • FDA Guidance: “Computer Software Assurance for Production and Quality System Software” (2022).
  • ISO 13485:2016, clauses 4.1.6 (software in QMS), 7.3 (design and development), 7.4 (purchasing) as applicable.
  • IEC 62304:2006+A1:2015 — software lifecycle (often implemented under design controls).

Changelog

  • 1.0.0 (2026-03-20): Initial skill covering 820.30, DHF, verification vs validation, ISO 13485 slice.
This skill doesn’t include stateful context yet, instructions only. Learn about stateful skills.

Security

Loading security scan...

Reviews (0)

Browse all →
meddev-authenticationDefine secure user/device authentication, session management, and RBAC suitable for medical device risk profiles. Use when working on authentication in medical device software.0 installsmeddev-state-machinesDefine explicit, testable state machines with safe states, controlled transitions, and error handling suited to medical device safety classes. Use when working on state machines in medical device software.0 installsmeddev-change-controlDefine controlled change-request workflow for software: impact analysis, approvals, verification scope, and regulatory submission flags. Configuration identification and baselines are in DOC-SCM (IEC 62304 Clause 8.1, 8.3); approved implementation and re-verification align with 8.2. Use when working on change control in medical device software.0 installsCONN-USBImplement safe USB device connectivity: class selection, enumeration handling, data validation, electrical and safety considerations.0 installsmeddev-ai-ml-medical-device-softwareAddress machine learning inside the regulated product (not LLM coding assistants): Good Machine Learning Practice (GMLP), training/validation data integrity, Predetermined Change Control Plans (PCCP) where applicable, drift monitoring, and linking models to ISO 14971 and software lifecycle controls. Use when working on ai ml medical device software in medical device software.0 installsTEST-HILValidate software with real or representative hardware, sensors, actuators, and timing, to uncover integration and safety issues not visible in simulation.0 installs

More from Medical Device Skills

View source →

More Build skills

Browse category →
senior-frontendBuild and review React, Next.js, TypeScript or Tailwind frontends: components, scaffolding, performance and bundle size. Use whenever the user wants a new component or app, a page feels slow, the bundle is too big, or asks for a frontend code review.647 installsui-design-systemUse when building a design system: tokens, component docs, responsive scales or developer handoff. Generates tokens with design_token_generator.py.549 installsfrontend-designGuidance for distinctive, intentional visual design when building new UI or reshaping an existing one. Helps with aesthetic direction, typography, and making choices that don't read as templated defaults.358 installssenior-backendBuild and fix backend services: REST or GraphQL APIs, auth, microservices, PostgreSQL queries, migrations and load tests. Use whenever the user's API is slow or broken, needs login or auth added, a schema change, or a new endpoint or service scaffolded.259 installsself-improving-agentCurate what the agent has learned into durable project knowledge. Reviews Claude Code auto-memory (MEMORY.md and its topic files) for learnings worth keeping, promotes proven ones into CLAUDE.md or .claude/rules/, records a correction as a lesson that persists across sessions, prunes stale or duplicate entries, and turns a recurring fix into a reusable skill. Use whenever the user asks what you have learned or saved about the project, which memories should become permanent rules, to remember a lesson or stop repeating a mistake, or to clean up, audit or check memory health.183 installsusing-superpowersUse when starting any conversation - establishes how to find and use skills, requiring skill invocation before ANY response including clarifying questions147 installs

Frequently asked questions about REG-FDA820

What does the REG-FDA820 skill do?

Map **21 CFR Part 820 Subpart C — Design Controls** to software lifecycle work: design planning, inputs, outputs, review, verification, validation, transfer, and DHF/DMR links, including how **software validation** (intended use / risk controls) complements **verification** (requirements and specs). Includes a concise **ISO 13485:2016** slice where it overlaps QMS expectations for design and risk (interpretation; align with your notified-body MDSAP scope for EU). It's a reusable SKILL.md instruction set that loads into your AI coding assistant on demand, no prompt engineering, no copy-pasting every session.

How do I install the REG-FDA820 skill?

Run `npx @skills-hub-ai/cli install meddev-skills-reg-fda820` from your terminal. The CLI writes the SKILL.md to the correct location for your AI tool (e.g. ~/.claude/skills/meddev-skills-reg-fda820/ for Claude Code or ~/.cursor/skills/ for Cursor with --target cursor) and adds it to your project's .skills.json lockfile.

Which AI tools does REG-FDA820 work with?

REG-FDA820 runs in Claude Code. It follows the open Agent Skills standard (SKILL.md), so the same skill works in every supported tool without modification.

Is the REG-FDA820 skill free?

Yes. Every skill on skills-hub.ai is free and open-source. There are no premium tiers, paywalls, or usage limits. You only pay for whatever AI assistant you're already using.

How do I use REG-FDA820 after installing it?

In Claude Code, type `/meddev-skills-reg-fda820` (or whatever slash command the skill registers) and the AI follows the skill's instructions immediately. You can also reference it by name in natural language, your AI loads the skill into context when relevant.

Can I share the REG-FDA820 skill with my team?

Yes. Commit your project's .skills.json lockfile and teammates run `npx @skills-hub-ai/cli install` (no args) to install every skill at the exact version you pinned. Organization-scoped installs work via skills-hub.ai organizations.